RAPIDPoint 400/405 Systems

Point-of-care efficiency with central lab accuracy

RAPIDPoint 400/405 Systems
 
Contact Us
Sample Size  
  Syringe: 200 µL
  Capillary: 100 µL

 

Analysis

  
Time to Results: Approximately 60 seconds

 

QC/AQC

  
Analysis Time 

130 seconds (automatic QC or ampule)

 

 

Measurement Cartridge

Use life: 28 days
Number of tests: 

250, 400, 750 

 

Calibration

  
1 Point Calibration: Every 30 minutes
2 Point Calibration: 

Every 2 hours 

 

System Dimensions 

Width: 30.0 cm. (11.5 in.)
Depth: 42.0 cm. (16.0 in.)
Height: 55.0 cm. (21.5 in., display at highest position)
Weight: 

15.5 kg. (34.0 lbs., excluding cartridges) 

 

External Interfaces 

  

RS-232, 10BaseT Ethernet port, Bar-code scanner port 

 

Power Requirements

Rating: 150 VA
Voltage: 100 to 240 VAC; (85 to 264 VAC autoranging)
Frequency: 

48 to 62 Hz 

 

Environmental

  
Temperature: 15ºC to 32ºC
Humidity: 5% to 85% non-condensing
Barometric Pressure:             

523 to 800 mmHg  

 

Approvals:

  
  UL Listed, CSA, IEC 601-1, JIS, FCC47-15-B, EN55011 (G1-1-B), EN60601-1-2

 

Related Products, Services & Resources